We can normally fulfill our respected consumers with our great excellent, great value and good provider due to we're much more specialist and extra hard-working and do it in cost-effective way for Gauge, Vacuum Pump Oil Filter, Vacuum Pump Ac System, "Making the Products of High Quality" is the eternal goal of our company. We make unremitting efforts to realize the goal of "We Will Always Keep in Pace with the Time".
Factory Supply Helium Leak Detection Services - Compound Vacuum Gauge, ZDF-62B5, 10E5 to 10E-7 Pa, 6 loops, Rs485 – KYKY Detail:
Model ZDF-62B5 compound vacuum gauge is composed of 2 sets of low vacuum measuring units and 1 set of high vacuum measuring units. One unit of resistance gauge (RG 2)is used to achieve the measurement (1.0×10E5Pa~1.0×10E-7Pa)independently, the unit of resistance gauge( RG1) combined with ionization gauge, as compound unit, is to achieve the continuous control and measurement (1×10E5~1×10E-7Pa).
Advantages:
1. Manual or automatic mode.
2. Convenient and flexible control
3. Long up-time, stability, reliability
4. Strong anti-jam capability ( preferred for R&D and industrial production)
Specifications:
| Compound Vacuum Gauge |
ZDF-62B5 |
| Measured loop |
3(resistance gauge: 2 , ionization gauge:1) |
| Type of Gauge |
ZJ-52T resistance gauge 2 sets
ZJ-14 ionization gauge 1 set |
| Measured range |
1.0×105Pa~ 1.0×10-7 Pa |
| Controlled range |
1.0×105Pa~ 1.0×10-7Pa |
| Controlled loop |
6 |
| Control mode: |
On or off relay control, Control point load: AC220V/3A DC28V/10A (Non-inductive load), Jog or regional control/with power off memory |
| Accuracy |
±1% |
| Power supply |
AC 220V±10% 50Hz |
| Sampling time |
1s |
| Power consumption |
45W |
| Interface |
Rs485 ( Modbus-RTU) |
| External control of Ionization gauge |
PLC |
| Display |
Five-digits LED display
e.g.: 1.2E-1 refers to 1.2×10-1Pa |
| Dimension(W*H*D) |
480*88*280 mm |
| Weight |
4Kg |
Product detail pictures:
Related Product Guide:
We support our purchasers with ideal high-quality merchandise and significant level company. Becoming the specialist manufacturer in this sector, we now have received loaded practical encounter in producing and managing for Factory Supply Helium Leak Detection Services - Compound Vacuum Gauge, ZDF-62B5, 10E5 to 10E-7 Pa, 6 loops, Rs485 – KYKY , The product will supply to all over the world, such as: Rotterdam, Uganda, Angola, We warmly welcome domestic and overseas customers to visit our company and have business talk. Our company always insists on the principle of "good quality, reasonable price, the first-class service". We have been willing to build long-term, friendly and mutually beneficial cooperation with you.