The company upholds the philosophy of "Be No.1 in quality, be rooted on credit and trustworthiness for growth", will continue to serve old and new customers from home and overseas whole-heatedly for Gate Valve Stainless, Valve Components, Becker Vacuum Pump Vanes, "Making the Products of High Quality" is the eternal goal of our company. We make unremitting efforts to realize the goal of "We Will Always Keep in Pace with the Time".
Low MOQ for Nash Liquid Ring Vacuum Pump - Resistance Vacuum Gauge, ZDR-12/06B, 1E-1 to 1E-8 Pa, Analog signal, 0-5V, Rs485 – KYKY Detail:
Model ZDR-12/06B vacuum gauge can achieve continuous measurement and control for 1*E-1~1*E-8Pa, and have analog signal output( 0-5V). Interface RS485. This ZDR-12/06B, has 1 measure loop and 2 control loops, the control accuracy can reach ±1%.
ZDR-12/06B is panel amount type with standard amounting dimensions.
Advantages:
1. convenient and flexible control
2. Long up-time, stability, reliability
3. Strong anti-jam capability ( preferred for R&D and industrial production)
Specifications:
| Resistance Vacuum Gauge |
|
| Measured loop |
1 |
| Type of Gauge |
ZJ-12 Ionization gauge |
| Measured range |
1.0×10-1Pa~1.0×10-8 Pa |
| Controlled range |
1.0×10-1Pa~1.0×10-8 Pa |
| Controlled loop |
2 |
| Control mode |
On or off relay control, Control point load: AC220V/3A DC28V/10A (non-inductive load), Jog or regional control/with power off memory |
| Accuracy: |
±1% |
| Analog output |
(0~5)V |
| Interface |
RS-48 |
| Sampling interval |
1s |
| Power consumption |
45W |
| Display |
Five-digits LED display, Scientific notation adopted, e.g.: 1.2E-1 refers to 1.2×10-1Pa |
| Operation temperature |
0℃~45℃ |
| Humidity |
≦85% |
| Power supply |
AC 220V ±10 , 50Hz |
| Install Dimensions(W*H*D) |
240*88*330mm |
| Weight |
4 Kg |
Product detail pictures:
Related Product Guide:
Well-run gear, qualified revenue workforce, and superior after-sales companies; We've been also a unified huge loved ones, anyone persist with the organization benefit "unification, determination, tolerance" for Low MOQ for Nash Liquid Ring Vacuum Pump - Resistance Vacuum Gauge, ZDR-12/06B, 1E-1 to 1E-8 Pa, Analog signal, 0-5V, Rs485 – KYKY , The product will supply to all over the world, such as: moldova, Tunisia, Cyprus, When you are keen on any of our goods following you view our product list, be sure to feel free to make contact with us for inquiries. You'll be able to send us emails and get in touch with us for consultation and we shall respond to you as soon as we're able to. If it's convenient, you could find out our address in our web site and come to our enterprise. or additional information of our products by yourself. We're generally ready to build lengthy and steady co-operation relations with any possible shoppers within the associated fields.