Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY

Short Description:


Product Detail

Product Tags

Related Video

Feedback (2)

We have a highly efficient team to deal with inquiries from customers. Our goal is "100% customer satisfaction by our product quality, price & our team service" and enjoy a good reputation among clients. With many factories, we can provide a wide range of Oil Sealed Rotary Vacuum Pump, Helium Sniffer Leak Detector, Compressor Vacuum Pump, Our group members are goal to provides merchandise with significant performance cost ratio to our consumers, as well as target for all of us is usually to satisfy our consumers from all around the environment.
Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY Detail:

The ZQJ-3200 Helium Leak Detectors are microprocessor-controller leak detecting instruments. All the processes in the instrument are controlled automatically.

Advanteges:

1. Easy operation–Small volume, light weight, compact structure, remote operation panel

2. Multiful interface–Rs232, Digital I/O, USD port

3. Powerful functions–Various test mode, capability of detecting H2. 3He, amd 4He, multiple menu setting

4. Reliable performance–High sensitivity, wide measurement range, high inlet pressure, fast response time

5. Reliable quality–Prolong the service life, resistance of yttrium oxide iridium filament

Specifications:

Type ZQJ-3200
Smallest Detectable Leak Rate(Pa•m3/s) 5×10-13  vacuum mode5×10-10  sniffing mode
Leak Rate Display(Pa•m3/s) 10-13~10-1
Maximum Inlet Pressure(Pa) 2500
Response Time(s) ≤2
Run-up Time(min) <3
Power 230 VAC±10%/50 Hz
120V±10%/60 Hz,10A
Working Temperature and Relative humidity Working Temperature 10~35℃,realtive humidity ≤80%
L*W*H(mm) 550×460×304
Weight(kg) 44

Methods:

Vacuum Method

In the vacuum method test gas is blown against the wall of the evacuated sample from the atmosphere side. It enters the sample at leaks and is fed to the leak detector.

The sample must be vacuum pressure-proof.

The sensitivity stages GROSS —FINE—ULTRA are run through.

The detection limit is lower than in the sniffing method. The helium concentration at the leak must be known in order to quantify the leak. The state of equilibrium must be waited for.

Sniffing Method

In the sniffing method the test gas escaping from leaks in the sample into the atmosphere is detected.

The sample must withstand the applied test pressure.

In operation with the sniffing probe a constant gas flow is sucked in from the atmosphere. The helium proportion of the air (5.2 ppm) causes a leak rate display of approx. 1*10-6 mbar l/s which can be eliminated by the ZERO function.20 3 Description Operating Instructions, ikna88en1-01, 1605

To detect a leak, the sniffing probe is applied to the points of the sample under helium overpressure which are suspected of leaking. An increased leak rate value indicates an increased concentration of helium and therefore a leak. The higher the pressure and the helium concentration in the sample, the smaller the leaks which can be detected.

The sensitivity stages GROSS—FINE are run through.

The detection sensitivity and the quantifiability of the leak rate are less favourable than in the vacuum pressure leak detection.



Product detail pictures:

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures

Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY detail pictures


Related Product Guide:

Our products are widely recognized and trusted by users and can meet continuously changing economic and social needs for Manufacturer for Helium Vacuum Leak Detector - Helium Leak Detector, ZQJ-3200, Min rate 5*1E-13, Display 5E-13 to 1E-1 – KYKY , The product will supply to all over the world, such as: Juventus, Boston, Lithuania, Due to the changing trends in this field, we involve ourselves into products trade with dedicated efforts and managerial excellence. We maintain timely delivery schedules, innovative designs, quality and transparency for our customers. Our moto is to deliver quality products within stipulated time.
  • Hope that the company could stick to the enterprise spirit of "Quality, Efficiency, Innovation and Integrity", it will be better and better in the future.
    5 Stars By Diana from Mauritius - 2018.06.18 19:26
    The factory workers have rich industry knowledge and operational experience, we learned a lot in working with them,we are extremely grateful that we can encount a good company has excellent wokers.
    5 Stars By Phyllis from Boston - 2017.11.01 17:04
    Write your message here and send it to us